TRUMPF Medical Systems Recalls Television Mounting System Due to Falling Hazard
What was recalled
The mounting systems were sold for use in hospital Intensive Care Units. The units are off-white and have multiple joints between the ceiling mount and the TV mount at the end of the arm system.
Hazard
The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
Injuries
The firm received one report of a mounting system falling while being positioned. No injuries have been reported.
Made in
Germany
Where it was sold
- TRUMPF Medical Systems Inc. nationwide From January 2005 through December 2011 for about $2,100.
Photos

Shopping for a replacement? Safer certified picks
If your item was recalled, these brands hold independent safety certifications and have no recall on record.
Avocado Green Mattress
Certified organic latex, wool, and cotton — no fiberglass.
Check pricePlushBeds
Natural latex, multiple whole-product certifications.
Check priceCertifications shown are editorial references — verify current status with the brand. Some links are affiliate links; we may earn a commission at no cost to you. See our disclosure.
Other fall / collapse recalls

Gazelle USA Recalls Urban Arrow Cargo Electric Bikes Due to Fall Hazard
Gazelle USA

Sauna360 Expands Recall of Tylö Halmstad and Kiruna Hybrid Saunas Due to Fall Hazard; New Remedy Provided
SFactor

MONDRAKER USA Recalls Mountain Bicycles Equipped with ONOFF S9 Carbon MTB Handlebars Due to Risk of Serious Injury or Death from Fall Hazard
MONDRAKER USA

Branch Recalls Ergonomic Chairs Due to Fall Hazard
Branch